Computed Tomography Radiation Doses Vary Substantially Across Two Tertiary

A retrospective evaluation of 6,451 clinical computed tomography examinations performed across two tertiary hospitals in Taif, Saudi Arabia, has established that scanner radiation output and overall examination exposures differ markedly between medical centers

Illustrative imageIllustrative image

A retrospective evaluation of 6,451 clinical computed tomography examinations performed across two tertiary hospitals in Taif, Saudi Arabia, has established that scanner radiation output and overall examination exposures differ markedly between medical centers operating within the same metropolitan area. The scientific investigation, conducted using patient records gathered from King Faisal Medical Complex and King Abdulaziz Specialist Hospital between January 2024 and December 2025, demonstrated that radiation indicators for standard imaging procedures fluctuated significantly across anatomical scan regions, patient age groups, and institutional facilities. Published evidence underscores that computed tomography remains an indispensable diagnostic modality in modern medicine owing to its rapid acquisition times and high-resolution cross-sectional visualization of complex internal structures. However, while computed tomography procedures constitute a relatively modest percentage of aggregate medical imaging exams, their increasing utilization worldwide has driven cumulative patient exposure higher. This upward trend in medical exposure has heightened institutional scrutiny surrounding potential radiation risks, reinforcing the fundamental radiological objective of optimizing acquisition techniques to preserve essential diagnostic image quality while keeping radiation exposure as low as reasonably achievable.

Radiation protection requirements are especially pressing within pediatric imaging departments because growing bodies exhibit heightened radiosensitivity compared to adults. Infants and young children possess anatomical organs and physiological tissues characterized by larger fractions of actively dividing cellular populations, rendering biological structures substantially more vulnerable to ionizing radiation damage. Furthermore, younger patients enjoy significantly longer statistical life expectancies, which inherently expands the temporal window for late stochastic complications, notably the lifetime risk of developing radiation-induced malignancies, to manifest following exposure. To counter these vulnerabilities, international radiological health guidelines consistently recommend the implementation of dedicated, pediatric-specific imaging protocols tailored to body size and chronological age rather than relying on crude modifications of adult settings. Radiation dose optimization is formally anchored by the concept of diagnostic reference levels, defined by the International Commission on Radiological Protection not as immutable regulatory ceilings or statutory dose caps, but rather as practical benchmarking tools designed to identify unusually elevated or depressed radiation doses during standard examinations so clinical teams can adjust techniques accordingly.

The retrospective cohort under study drew upon clinical imaging archives encompassing two major public healthcare facilities serving the Taif population, King Faisal Medical Complex and King Abdulaziz Specialist Hospital. Investigators retrieved digital records spanning the two-year period from January 2024 through December 2025 between January and March 2026, sourcing data directly from institutional Radiology Information Systems and Picture Archiving and Communication Systems. In total, the study population comprised 6,451 individual computed tomography examinations, treating each unique scan as the primary unit of statistical analysis rather than tracking individual patient identities. The vast majority of the documented procedures involved patients aged 15 years and older, who collectively accounted for 5,901 adult examinations, or approximately 91.5 percent of the analyzed dataset. The remaining 550 examinations, representing 8.5 percent of the total sample, were performed on pediatric individuals under the age of 15. The threshold of 15 years was derived directly from the operational categorization historically maintained by the participating facilities, meaning adolescent patients aged 15 to 17 were grouped into the adult evaluation stratum.

Technical equipment across the two tertiary institutions involved five distinct multi-detector computed tomography scanners, with two operational systems situated at King Faisal Medical Complex and three installed at King Abdulaziz Specialist Hospital. Institutional records confirmed that standard quality control evaluations were routinely performed across all five units, with recent documented testing dates spanning late 2025 through early 2026 across various equipment platforms from General Electric HealthCare, Philips Healthcare, and Siemens Healthineers. Specifically, King Abdulaziz Specialist Hospital operated a 2014 Philips scanner inspected in October 2025, a 2019 GE scanner assessed in February 2026, and a 2016 Siemens scanner checked in January 2026. King Faisal Medical Complex operated two General Electric units, consisting of a Discovery CT 750 HD manufactured in 2012 and evaluated in March 2026, alongside a Revolution Discovery system manufactured in 2018 and inspected in February 2026. However, examination-specific acquisition settings—including tube voltage, current, pitch, automated exposure controls, iterative reconstruction parameters, and scan phase counts—were recorded inconsistently across the legacy records and could not be evaluated.

To evaluate patient radiation exposure consistently across scanner models and clinical departments, researchers collected two primary standard dosimetric indices: the volume computed tomography dose index and the dose–length product. In contrast, the dose–length product reflects the total cumulative radiation emitted over the complete spatial length of the anatomical examination, combining scanner output intensity with total longitudinal scan coverage. For each sufficiently sampled patient group and scan category, the 75th percentile of the observed dose distribution was calculated to define local patient-level diagnostic reference levels, denoted as DRL75. Following standardized sampling guidance issued by the International Commission on Radiological Protection in Publication 135, local reference levels were formally reported only for demographic and anatomical strata containing a minimum threshold of 20 completed examinations.

Analysis of pediatric computed tomography examinations demonstrated substantial variation depending on anatomical scan location, patient age brackets, and the treating medical facility. Although the global study cohort contained 550 pediatric procedures, detailed dosimetric calculations were conducted on 544 examinations after six records from King Faisal Medical Complex were omitted due to missing dosimetric data fields. Head examinations consistently registered the highest scanner output and overall radiation burden across every assessed pediatric cohort, substantially exceeding values recorded for thoracic and abdominal scans. At King Faisal Medical Complex, pediatric head volume dose index values reached a mean of 39.97 milligrays in infants under one year, 38.21 milligrays in the one-to-five-year category, 43.97 milligrays in children aged five to ten, and 46.01 milligrays in the ten-to-fifteen-year cohort. For pediatric patients aged 10 to 15 years undergoing head scans, the derived DRL75 at King Faisal Medical Complex was 53.58 milligrays for volume dose index and 934.50 milligray-centimeters for dose–length product, whereas King Abdulaziz Specialist Hospital exhibited notably lower benchmarks of 34.13 milligrays and 650.50 milligray-centimeters respectively.

Among adult patients aged 15 years and older, radiation measurements likewise displayed prominent differences across anatomical regions and institutional settings. Head imaging accounted for the highest absolute volume computed tomography dose index values at both medical centers, whereas thoracic and abdominal examinations exhibited greater relative heterogeneity between the two clinical environments. At King Faisal Medical Complex, the adult head examination cohort encompassed 3,418 scans, yielding a mean volume dose index of 44.92 milligrays and a 75th percentile benchmark of 51.22 milligrays, with an accompanying dose–length product DRL75 of 982.53 milligray-centimeters. In comparison, King Abdulaziz Specialist Hospital recorded 82 adult head procedures, producing a mean volume dose index of 36.88 milligrays, a DRL75 of 44.46 milligrays, and an associated dose–length product benchmark of 831.00 milligray-centimeters. Adult chest imaging demonstrated even sharper contrasts: King Faisal Medical Complex recorded a chest volume dose index DRL75 of 30.52 milligrays and a dose–length product DRL75 of 548.73 milligray-centimeters across 1,482 procedures, compared to 10.15 milligrays and 353.70 milligray-centimeters across 31 chest examinations evaluated at King Abdulaziz Specialist Hospital.

Direct descriptive comparisons between King Faisal Medical Complex and King Abdulaziz Specialist Hospital revealed that King Faisal Medical Complex registered higher 75th percentile volume dose metrics across all three investigated anatomical regions. For adult head scans, the absolute difference in volume dose index DRL75 was 6.76 milligrays, representing a 15.2 percent higher figure at King Faisal Medical Complex, alongside an 18.2 percent higher dose–length product benchmark. In adult thoracic examinations, the institutional difference expanded dramatically to a 20.37 milligray gap in volume dose index, meaning the benchmark at King Faisal Medical Complex was 200.7 percent higher than that of King Abdulaziz Specialist Hospital, accompanied by a 55.1 percent difference in dose–length product. In adult abdominal scans, King Faisal Medical Complex reported an index DRL75 of 35.18 milligrays compared to 14.56 milligrays at King Abdulaziz Specialist Hospital, reflecting a 141.6 percent divergence, although abdominal dose–length products differed by only 4.0 percent. The researchers highlighted that these comparisons represent purely descriptive calculations rather than inferential proof of statistical divergence, particularly given the severe procedural volume disparity between the two hospitals.

When the newly calculated local reference levels were benchmarked against published national and international reference standards, findings diverged based on procedure type and anatomical region. Conversely, adult thoracic and abdominal volume dose index reference levels at King Faisal Medical Complex exceeded external baselines, with its chest DRL75 of 30.52 milligrays standing markedly above the Saudi national benchmark of 12 milligrays and the United Kingdom figure of 8.5 milligrays. For pediatric cranial examinations, benchmarks at King Faisal Medical Complex were higher than national reference figures in older children, with its ten-to-fifteen-year head index benchmark of 53.58 milligrays surpassing both the Saudi national reference level of 42 milligrays and the British standard of 46 milligrays, whereas King Abdulaziz Specialist Hospital achieved lower pediatric cranial benchmarks of 34.13 milligrays.

The study authors emphasized several critical methodological limitations that must guide the clinical interpretation of these radiological findings. Because patient size parameters—such as recorded body weight, body mass index, and cross-sectional effective diameter—were unavailable in the retrospective archives, researchers could not compute size-specific dose estimates. Furthermore, the absence of granular acquisition settings and reconstruction protocols prevented the research team from identifying the precise mechanical or protocol factors responsible for the observed dosage disparities between facilities. Substantial examination volume imbalances also existed, as King Faisal Medical Complex provided 6,183 examinations while King Abdulaziz Specialist Hospital contributed only 268 scans, leaving multiple pediatric subcategories with fewer than 20 cases and preventing formal DRL75 calculation for those cohorts. Consequently, the authors concluded that radiation dose optimization in computed tomography must operate as an institution-specific, age-tailored, and examination-focused ongoing process, relying on continuous internal surveillance, protocol standardization, and regular re-evaluation of local reference benchmarks to protect patient health without degrading diagnostic utility.

Subject of Research: Radiation dose distributions and diagnostic reference levels for computed tomography examinations in tertiary medical centers

Article Title: Establishment of local diagnostic reference levels for adult and pediatric computed tomography examinations in two tertiary hospitals in Taif, Saudi Arabia

Article References: Osman, H., Khandaker, M. U., Alharthi, M., Alotaibi, R., Alwathnani, M., Althagafi, B., Alotaibi, S., Metebi, A. M., Asiri, M., & Algarni, A. F. (2026). Establishment of local diagnostic reference levels for adult and pediatric computed tomography examinations in two tertiary hospitals in Taif, Saudi Arabia. Journal of Radiation Research and Applied Sciences, 19(4), Article 102824. https://doi.org/10.1016/j.jrras.2026.102824

Image Credits: AI Generated

DOI: 10.1016/j.jrras.2026.102824

Keywords: computed tomography, diagnostic reference levels, radiation dosage, CTDIvol, dose length product, pediatric radiology, adult radiology, radiation protection, King Faisal Medical Complex, King Abdulaziz Specialist Hospital, Taif healthcare, medical imaging optimization, ALARA principle, dosimetry, radiological quality assurance, Saudi healthcare studies

News Source: Ophelia Keating. (November 24, 2026). Computed Tomography Radiation Doses Vary Substantially Across Two Tertiary Hospitals in Taif. Scienmag.